Question regarding OptiTrack integration and transform processing in Plus Toolkit #1272
We are currently working on real-time 3D ultrasound reconstruction using motion capture data streamed to PlusServer and 3D Slicer.
For Vicon and QTM systems, we currently use a custom Python bridge:
Mocap SDK → OpenIGTLink server (pyigtl) → Plus OpenIGTLinkTracker device
In this configuration, we observe slight rigid body jitter (especially rotational jitter with Vicon/Nexus Kinematic Fit), which produces visible waviness artifacts in our reconstructed 3D ultrasound volumes.
Interestingly, similar experiments performed with OptiTrack appear significantly more stable.
We were therefore wondering whether the OptiTrack integration in Plus Toolkit performs any additional processing internally, such as:
temporal smoothing
interpolation
buffering
transform filtering
timestamp correction
or any other stabilization step
compared to a generic OpenIGTLinkTracker input such as the one we currently use.
In other words:
does the OptiTrack integration in Plus simply forward rigid body transforms,
or is there additional processing involved that could explain improved stability?
We are trying to determine whether the observed differences come purely from the mocap systems themselves, or whether part of the stability may also originate from the Plus integration pipeline.
Thank you very much for your help and clarification.
It would appear that Plus does simple retrieval of the transform from the OptiTrack. There is an option for a Motive configuration file, so perhaps that might tell the OptiTrack to perform smoothing.
There are a number of other tracking devices that perform on-board data processing, but for the most part, Plus simply retrieves the transforms that the trackers send out.
Plus does do temporal interpolation to synchronize a video frame with an interpolated tracking frame.
Yes, we were using OptiTrack directly through the native Plus integration.
Thank you very much for the clarification — this is what we suspected, but we wanted to make sure there was no additional hidden processing on the Plus side.
We are aware of the temporal calibration/interpolation performed in Plus, and we are also able to achieve similar temporal calibration values with Nexus (around 150 ms offset).
We also checked the MotiveProfileFile configuration and could not find any smoothing-related parameters there; it seems to mainly contain camera and rigid body configuration information.
What is puzzling for us is that I have tested several other tracking systems in very similar 3D ultrasound reconstruction workflows (OptiTrack Flex13, OptiTrack Duo, NDI Aurora bars, etc.) and I never observed the same fluidity/reconstruction artifacts that we currently see with Nexus.
So at this stage we believe the issue is probably related either to the Vicon tracking/reconstruction itself or to how the transforms are produced upstream, and we are continuing to investigate the origin of the problem.
Activity
adamrankin commentedon Apr 29, 2026
It would appear that Plus does simple retrieval of the transform from the OptiTrack. There is an option for a Motive configuration file, so perhaps that might tell the OptiTrack to perform smoothing.
There are a number of other tracking devices that perform on-board data processing, but for the most part, Plus simply retrieves the transforms that the trackers send out.
Plus does do temporal interpolation to synchronize a video frame with an interpolated tracking frame.
Were you using the OptiTrack with Plus?
AurelieSar commentedon Apr 30, 2026
Yes, we were using OptiTrack directly through the native Plus integration.
Thank you very much for the clarification — this is what we suspected, but we wanted to make sure there was no additional hidden processing on the Plus side.
We are aware of the temporal calibration/interpolation performed in Plus, and we are also able to achieve similar temporal calibration values with Nexus (around 150 ms offset).
We also checked the MotiveProfileFile configuration and could not find any smoothing-related parameters there; it seems to mainly contain camera and rigid body configuration information.
What is puzzling for us is that I have tested several other tracking systems in very similar 3D ultrasound reconstruction workflows (OptiTrack Flex13, OptiTrack Duo, NDI Aurora bars, etc.) and I never observed the same fluidity/reconstruction artifacts that we currently see with Nexus.
So at this stage we believe the issue is probably related either to the Vicon tracking/reconstruction itself or to how the transforms are produced upstream, and we are continuing to investigate the origin of the problem.
Thank you for your quick answer,
Best
Aurélie